AFX SERIES® OPERATION MANUAL
SECTION 8:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 244 of 485
8.6.1.2
Transient Segment Output Parameter List Table
The following table details the available parameters for the transient segment program
definitions.
Parameters
Description
SEGment,<n>
Transient segment n of the selected program to be edited. Also the
starting segment when querying multiple segments (see NSEGS)
n = <1> to <100>
FSEG,<n>
sets objective Frequency of selected segment
n = <:SOUR:FREQ:LIM:MIN> to <:SOUR:FREQ:LIM:MAX> Hz
VSEG,<n>
sets objective voltage ФA, ФB, ФC of selected segment
n = <0> to <300 x XFMRRATIO> Volts
VSEG1,<n>
sets objective voltage ФA of selected segment
n = <0> to <300 x XFMRRATIO> Volts
VSEG2,<n>
sets
objective voltage ФB of selected segment
n = <0> to <300 x XFMRRATIO> Volts
VSEG3,<n>
sets objective voltage ФC of selected segment
n = <0> to <300 x XFMRRATIO> Volts
WFSEG,<n>
sets Waveform ФA, ФB, ФC of selected segment
n = <1> to <16>
WFSEG1,<n>
sets
Waveform ФA of selected segment
n = <1> to <16>
WFSEG2,<n>
sets Waveform ФB of selected segment
n = <1> to <16>
WFSEG3,<n>
sets Waveform ФC of selected segment
n = <1> to <16>
TSEG,<n>
sets execution time (to reach objective Voltage and Frequency) of
selected segment
n = <0> or <0.0002> to <300> seconds. Setting n to a negative value
such as -5 selects cycle-based transient operation, i.e., each
segment time is equal to the period of 5 cycles (1 cycle = 1/FREQ
secs)
LAST
sets selected segment to be the last segment of selected transient
Query: The NSEGS parameter of a :PROG:DEF? query returns total
number of Transient segments defined in the selected program
Содержание AFX Series
Страница 484: ......