AFX Series Power Source Operation Manual
Page 261 of 485
Command Syntax
[SOURce:]CONFIG:DC <b>
Description
Selects alternative DC loop compensation mode for improved DC
transient response.
The effect of this command depends on the selected output mode as
follows:
AC Mode:
No effect
DC Mode:
Changes the DC loop compensation
AC+DC Mode: No effect
Note:
Available in units with Firmware revision 1.6.0 or higher.
Parameters
< 0 | 1 >
( 0 = Normal loop, 1 = Faster Loop )
Parameter Format
<b>
Example
SOUR:CONFIG:DC 1
Query Format
[SOURce:]CONFIG:DC?
Returned Data Format
<b>
Query Example
CONFIG:DC?
0
Command Syntax
[SOURce:]CONFIG:HFreq <b>
Description
This command enables high-frequency output current protection
extension mode. In this mode, the protection for high frequency
content at the output of the power source is held off for up to 2
seconds to allow short time events to ride through without tripping
the normal protection mode.
Note:
Available in units with Firmware revision 1.6.6 or higher.
Parameters
< 0 | OFF | 1 | ON >
( 0 = Off, 1 = On)
Parameter Format
<b>
Example
SOUR:CONFIG:HF 1
Query Format
[SOURce:]CONFIG:HFreq?
Returned Data Format
<b>
Query Example
CONFIG:HF?
1