AFX SERIES® OPERATION MANUAL
SECTION 8:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 255 of 485
Command Syntax
PROGram:TRANsient:MODE
Description
This commands selected between legacy UPC controller mode
SEGMENT transient mode or STEP mode. Segment mode is backward
compatible with PPS UPC controllers and use a steady state table and
a transient table for each segment. A total of 99 segments can be
programmed. STEP mode uses the conventional SCPI LIST system of a
list of transient steps executed sequentially.
Parameters
< 0 | SEGMENT | 1 | STEP >
Parameter Format
<cr>
Example
PROG:TRAN:MODE STEP
Query Format
PROGram:TRANsient:MODE?
Returned Data Format
<cr>
Query Example
PROG:TRAN:MODE?
1
Command Syntax
PROGram:TRANsient:CSC
Description
This command turns the CSC Mode during transient execution on (1)
or off (0). This allows load regulation adjustment during transient
execution. However, keep in mind that CSC adjustments are based on
RMS voltage measurement, which take several cycles to run so
turning on CSC on fast changing voltage transients is not
recommended as the CSC may interfere with the programmed
transient voltages.
Note 1
: In UPC compatibility mode, the CSC is always off during
transient execution regardless of this setting to match the UPC
controller operation.
Note 2
: This command is available on units with firmware revision
1.3.0 or higher.
Parameters
< 0 | OFF | 1 | ON >
Parameter Format
<cr>
Example
PROG:TRAN:CSC ON
Query Format
PROGram:TRANsient:CSC?
Returned Data Format
<cr>
Query Example
PROG:TRAN:CSC?
1