Command Syntax
[SOURce:]FREQuency:LIMit:MINimum
Description
Sets the lower user limit for frequency programming. Not that the
lower limit set value must be less than the upper limit set value or an
error will be generated and no change to the set value will take place.
The lower limit set value must fall within the source specified
frequency range capability.
Parameters
1.00 – 1200.0
Parameter Format
<nr2>
Example
FREQ:LIM:MIN 47.0
Query Format
[SOURce:]FREQuency:LIMit:MINimum?
Returned Data Format
<nr2>
Query Example
FREQ:LIM:MIN?
47.0000
Command Syntax
[SOURce:]FREQuency:LIMit:MAXimum
Description
Sets the upper user limit for frequency programming. Note that the
upper limit set value must be greater than the lower limit set value or
an error will be generated and no change to the set value will take
place. The upper limit set value must fall within the source specified
frequency range capability.
Parameters
15.00 – 3000.0
Parameter Format
<nr2>
Example
FREQ:LIM:MAX 63.0
Query Format
[SOURce:]FREQuency:LIMit:MAXimum?
Returned Data Format
<nr2>
Query Example
FREQ:LIM:MAX?
63.0000
Query Format
[SOURce:]FREQuency:LIMit:RANGe?
Description
This command returns the lower and upper frequency user limit set
value.s
Returned Data Format
<nr2>, <nr2>
Query Example
FREQ:LIM:RANG?
53.0000,63.0000
Command Syntax
[SOURce:]FREQuency:SLEW
Description
Sets the frequency slew rate in Hz/msec. This command is not valid
when in DC mode.
Parameters
0.01 – 1200.00
Parameter Format
<nr2>
Example
FREQ:SLEW 100.0
Query Format
[SOURce:]FREQuency:SLEW?
Returned Data Format
<nr2>
Query Example
FREQ:SLEW?
100.0000