AFX SERIES® OPERATION MANUAL
SECTION 8:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 305 of 485
Command Syntax
[SOURce:]STEP:REPeat <no, opt>
Description
This command sets the number of repeat times for the ramp of the
step
Parameters
No of repeats, Option: < NORmal | SATurate >
Default is NORmal if optional second parameter is omitted.
Parameter Format
<nr1>, <cr>
Example
STEP:MODE:REP 100
Query Format
[SOURce:]STEP:REP?
Returned Data Format
<nr2>
Query Example
STEP:REP?
100, NOR
Query Syntax
[SOURce:]STEP:REPeat:MINimum?
Description
This command returns lowest permissible set value for the STEP
length time setting.
Returned Data Format
<nr1>
Query Example
STEP:REP:MIN?
0
Query Syntax
[SOURce:]STEP:REPeat:MAXimum?
Description
This command returns highest permissible set value for the STEP
length time setting.
Returned Data Format
<nr1>
Query Example
STEP:REP:MAX?
65535
Query Syntax
[SOURce:]STEP:REPeat:DEFault?
Description
This command returns default set value for the STEP length time
setting.
Returned Data Format
<nr2>
Query Example
STEP:REP:DEF?
1
Command Syntax
[SOURce:]STEP:COUNT <no, opt>
Description
This command sets the number of steps count.
Parameters
No of step count, Option: < NORmal | SATurate >
Default is NORmal if optional second parameter is omitted.
Parameter Format
<nr1>, <cr>
Example
STEP:MODE:REP 12
Query Format
[SOURce:]STEP:COUNT?
Returned Data Format
<nr2>
Query Example
STEP:REP?
12, NOR