Single Camera Alignment
The alignment of a single camera against a reference comprises first the capture the chosen test
chart/scene by the camera to which you want to match other cameras, followed by the capture of
the same scene by the camera(s) under test.
1. Capturing the Reference Image
The reference image may be captured using the
Capture Reference
option on the Button Bar of
the 3D Meters display. This captures the current frame from Input 1 and adds it to the Gen
window.
Note:
The reference image can be saved and restored to the Generator window in the same way
as any other captured frame: it is only the name that is special.
Control Panel
Mouse & Keyboard
To capture a Reference image:
1.
Set up your ‘Reference’ camera in required fashion and focus on the Reference Test Chart
2.
Feed the input from the Reference camera into the port currently selected as Input 1.
3.
Call up either the 3D Meters display (as described above) or the Chroma Sabres display (see
Section I.9).
4.
Select
Capture Reference
button on Button Bar.
The current frame image is then captured and displayed on the Generator window as the pattern
Reference.yuv
.
To save the image for subsequent use:
Use the
Save
option from the Generator window’s Session menu to save the
Reference.yuv
pattern to your chosen location. Do NOT change the filename. (See Section Q.3 for further
details.)
2. Aligning the Test Camera
The following steps may be used to determine differences in set-up between a single ‘Test’
camera and the camera used when capturing the Reference image.
Control Panel
Mouse & Keyboard
To align single camera:
1.
Either:
Capture a reference image as described above
Or:
(i) Call up the
System
page of the Config window and set the
Analyser 2 Source
to
either the main Generator or Generator 2 (if fitted).
(ii) Use the Generator window’s
Load image
option to load the saved
Reference.yuv
(if
needed), then run this on the selected Generator (See Sections Q.1, Q.5 and Q.12 for details.)
2.
Focus the ‘Test’ camera on the scene used when capturing the reference image.
3.
Feed the input from the Test camera into SDI Input 1.
4.
Call up the
3D Meters
display.
5.
Set the
Scene Type
option to
Test chart
or
3D scene
as required.
6.
Set the
Alignment Source
to
In1 vs Ref
.
The displays in the window will then show how the input from the Test camera compares
against the selected Reference image.
7.
Adjust the set-up of the test camera to bring the values shown to zero or 1 as appropriate.
Содержание OTR 1001
Страница 1: ...Advanced Measurement Technology OTR 1001 User Guide Software Version 3 1 October 2013...
Страница 28: ...OmniTek OTR 1001 User Guide v3 1a A 16...
Страница 52: ...OmniTek OTR 1001 User Guide v3 1 B 24...
Страница 96: ...OmniTek OTR 1001 User Guide v3 1a E 24...
Страница 116: ...OmniTek OTR 1001 User Guide v3 1 G 6...
Страница 124: ...OmniTek OTR 1001 User Guide v3 1 H 8...
Страница 130: ...OmniTek OTR 1001 User Guide v3 1 I 6 Figure I 3 Difference view Figure I 4 Anaglyph representation...
Страница 184: ...OmniTek OTR 1001 User Guide v3 1 J 22...
Страница 238: ...OmniTek OTR 1001 User Guide v3 1 M 14...
Страница 250: ...OmniTek OTR 1001 User Guide v3 1 N 12...
Страница 254: ...O 4 OmniTek OTR 1001 User Guide v3 1...
Страница 280: ...Q 14 OmniTek OTR 1001 User Guide v3 1 Figure Q 12 Video Index options within the Gen window Properties...
Страница 283: ...OmniTek OTR 1001 User Guide v3 1 Q 17 Figure Q 15 Save File settings within the Gen window Properties...
Страница 292: ...Q 26 OmniTek OTR 1001 User Guide v3 1...
Страница 326: ...OmniTek OTR 1001 User Guide v3 1a S 26...
Страница 358: ...OmniTek OTR1001 User Guide v3 1 U 12...
Страница 364: ...OmniTek OTR 1001 User Guide v3 1 V 6...
Страница 379: ...OmniTek OTR 1001 User Guide v3 1 I 15 Example XML Statements Example Result...
Страница 386: ...OmniTek OTR 1001 User Guide v3 1 II 6...
Страница 389: ...OmniTek OTR 1001 User Guide v3 1...