FACT ADC?
FACT:ADC?
It is command which checks ADC Sampling speed/
Return value
"5Hz"
5times/1sec sampling time
"20Hz"
20times/1sec sampling time
"50Hz"
50times/1sec sampling time
"100Hz"
100times/1sec sampling time
"300Hz"
300times/1sec sampling time
"1.3KHz"
1300times/1sec sampling time
1.3KHz
1300times/1sec sampling time
ex)fact:ADC?
return value "5Hz"
FACT:LOAD-DEF
It is command which reset Factory Mode setting value.
However, calibration range will not be intialized, User-memory range will be cleared.
ex)fact:load-def
Reset factory mode setting
6-8. System Commands
Power supply controlling commands.
SYSTem:BEEPer
It is command which cause alarm once.
ex) syst:beep
It will be alarmed.
SYSTem:ERRor?
It is command which checks error messages from power supply.
Error messages will be saved up to 10 fixed memory, from 11th error, the lastest error message
will be deleted. If users check the error message, the message will be deleted.
Return value - error number ,"message"
ex) syst:err?
return value -222 "Out of data"
ex) syst:err?
return value 222, Out of data
1. Read reference "7. Error Messages" for further information.
2. Use CLS command to clear all error messages. *RST commands will not clear all error messages.
SYSTem:VERSion?
Users are able to check power supply version
Note
Users are able to check power supply version.
Return "YYYY.Ver"
YYYY - Indicates the YEAR
Ver - Indicates the version
ex) syst:vers?
return value "2008.3"
SYSTem:REMote
Set-up Power supply remote control.
ex) syst:rem
- 65 -