3-13. Output ON/OFF
Function that supplies or shut-off sources from output terminal.
Therefore, user could shut-off not removing the load.
Follow the sequence below :
▌Power Switch ON
Check whether "**OUTPUT OFF**" is displayed after input Power.
▌It is shut-off setting. To allow print source, press OUTPUT ON/OFF Key.
▌To shut-off sources, press OUTPUT ON/OFF once more.
3-14. Error Message Display
If an error occurs about calibration, self-testing mode, SCPI program error, the messages
will be saved in fixed memory, up to 10 messages.
Read reference "7. Error Messages" for further information about error message.
▌Error memory are stack structured, and recent message will be accessed first.
▌Error memory are stack structured, and recent message will be accessed first.
▌They will be saved up to 10, from 11th message, the earliest messages will be deleted.
▌When user check message by pressing ERR Key, the messages will be deleted.
▌ERR Lamp will be lighted if there is an error.
ERROR Check
▌Power Switch ON
Ch
k
h th
"
OUTPUT OFF
" i di
l
d ft
i
t P
Check whether "**OUTPUT OFF**" is displayed after input Power.
▌To check ERROR, press ERR key for several seconds.
If there is no error, display will be same as below, and automatically return
to previous page.
LCD Display Description
NO ERROR
ERR Lamp will light-out.
If there is an error, display will be like as below.(Error Code)
LCD Display Description
ERROR NO, -200
LCD Display Description
O
O,
00
▌To check next ERROR, press ERROR Key.
If there is an error, error code will be displayed.
LCD Display Description
ERROR NO, -10
▌To check next ERROR, press ERROR Key.
- 35 -