6-5. Measure Commands
This command measures readback voltage and current of power supply. It does not require
DVM(Digital Volt Meter) or Ammeter to measure.
MEASure:VOLTage[:DC]?
This command measure power supply output voltage.
Return value "voltage"
)
ex) meas:volt?
return value "4.0000"
MEASure:CURRent[:DC]?
This command measure power supply output current.
Return value "current"
ex) meas:curr?
return value "1.0000"
6 6 C lib ti
C
d
6-6. Calibration Commands
This is command about Remote Calibration.
CAL:VOLT {voltage|MIN|MAX}
Command about Voltage Calibration.
Please mind sequence of Calibration.
Read reference '4-6 Calibration Using Remote Interface"
> voltage Measures MIN range calibration voltage and MAX range calibration voltage
Save voltage value.
> MIN
Operates low range voltage calibration.
> MAX
Operates high range voltage calibration.
ex) Operate voltage calibration sequence.
CAL:VOLT MIN
Operate Low value.
CAL:VOLT voltage
Send measured voltage to DVM.
CAL:VOLT MAX
Low value is done, operate High value.
CAL:VOLT voltage
Send measured voltage to DVM.
When voltage value is sent after MIN or MAX value, it will be
saved in Calibration range automatically.
CAL:CURR {current|MIN|MAX}
Command about current Calibration.
Please mind sequence of Calibration.
Read reference"4-6 Calibration Using Remote Interface"
Read reference 4 6 Calibration Using Remote Interface
> current Measure MIN Range calibration of current value, and MAX Range calibration
of current value. Save current value.
> MIN
Operate Low range current Calibration.
> MAX
Operate High range current Calibration.
ex) Operate current calibration sequence.
CAL:CURR MIN
Operate Low value.
CAL:CURR current
Send measure current to DAM
CAL:CURR current
Send measure current to DAM.
CAL:CURR MAX
Low value is done, operate High value.
CAL:CURR current
Send measured current to DAM.
When current value is sent after MIN of MAX value, it will be saved in
Calibration range automatically.
- 62 -