5 5 AUTO CURSOR MV
5-5. AUTO CURSOR MV
▌This is function which moves voltage/current changed value to minimum when there is no
operation in front panel.
▌It is safe when user make a mistake using encoder switch.
▌Function
> DISABLE
Auto cursor MV is disabled.
> ENABLE
Auto cursor MV is enabled.
AUTO CURSOR MV Operation
▌Power Switch ON while pressing FACTORY Key.
Holding
▌Power Switch ON
▌
When
"
1. USER-MEM CLEAR
" Message displayed, release FACTORY Key.
▌
Use cursor key to move on to
"
3. AUTO CURSOR MV
" Menu.
Release
▌
Use cursor key to move on to 3. AUTO CURSOR MV
Menu.
▌Press FACTORY Key once more to set-up.
LCD Display Description
DISABLE
▌
Use cursor key to move to "ENABLE" and check Enable Mode.
LCD Display Description
ENABLE
▌Press FACTORY Key once more to apply.
LCD Display Description
Done