▌Use encoder switch to input voltage value shown in DVM Meter to power supply.
For example, if it is 9.2973V, input as below.
LCD Display Description
V-HIGH 9.297V
▌When it is done, press CAL Key to save.
LCD Display Description
ADC DATA F100H
Readback value will be
displayed in HEX for 500ms
LCD Display Description
2.CAL-VOLT HIGH
Return to previous CAL page.
▌Voltage Calibration is done, will be applied from next power supply operation.
Current CALIBRATION Operation
▌Connect measuring instrument before Calibration.
>Connect Power supply (+) terminal to electronic load input terminal (+), and connect
power supply (-) to input terminal (-) current monitoring resistor, and opposite side lead
to power supply (-) output terminal.
>Connect DVM input terminal (+) to Electronic load Current Monitoring resistor,
input terminal (-) to opposite side of Current Monitoring resistor.
▌P
P
S it h ON
hil
i
CAL K
▌Press Power Switch ON while pressing CAL Key.
▌Power Switch ON
▌When self-testing mode is done, it is ok to release CAL Key when
"1.CAL-VOLT LOW" is displayed.
Holding Key
Release
▌Use cursor key to select "3.CAL-CURR LOW" and press CAL Key.
LCD Display Description
A-LOW 0.0092A
Current can be different.
▌Wait for voltage shown on DVM Meter to be settled.
Wait for 5 mins