Third There can be a trip when load source. When Battery charge & discharge occurs,
its level can be higher than setting value.
Above diagram shows how to prevent it with inserting diode with right current capacity.
Fourth If it is difficult to manage Tripping, There is a way to Turn OCP Function OFF in force.
To turn it OFF, go to Factory Mode, and "Disable" "6.OCP USE".
▌When OCP trip occurs, "OCP TRIPPED" message will be displayed.
▌To clear OCP trip, firstly raise OCP Level value.
Press PROTECT Key to set-up Level value.
LCD Display Description
PROTECT> 1.OVP
▌Use right cursor key to move on to OCP Menu.
LCD Display Description
PROTECT> 2.OCP
▌Press PROTECT Key once more in "PROTECT> 2.OCP" condition.
LCD Display Description
OCP SET> 3.3000A
▌Use cursor key to change the value.
▌Use encoder switch to change Level value higher
▌Use encoder switch to change Level value higher.
▌Press PROTECT Key to save.
LCD Display Description
Changed
Return to previous menu when above message is displayed.