3-8. Programming Under Current Limit(UCL)
Users are able to change 0A to maximum current using encoder switch. However, when users
input UCL to the current, from this point, UCL current to maximum current are able to change.
UCL Level Value below can be different.
▌Power Switch ON
Check whether "**OUTPUT OFF**" is displayed after input Power.
▌Before setting UCL, current limit should be higher than UCL setting value.
Therefore, start with current.
▌Use cursor key to move on to voltage range, and use encoder switch to
change the voltage value.
Voltage cannot rise more than the Protection Level in Protection (OVP/UVL
will be displayed and voltage will be set-up protection level value
/OVL) If it tries to rise more than protection level, "PROTECTION LIMIT"
will be displayed and voltage will be set up protection level value.
▌Press PROTECT Key to set-up UCL Level.
LCD Display Description
PROTECT> 1.OVP
▌Use right cursor key to move on to UCL Menu.
LCD Display Description
PROTECT> 5.UCL
▌Press PROTECT Key once more in "PROTECT> 5.UCL" condition.
LCD Display Description
UCL SET> 0.0000A
▌Use cursor key to change the value.
▌Rotate encoder switch to change level value.
To prevent user's mistake, UCL Level cannot be higher than
p
,
g
current setting voltage.
▌Press PROTECT Key to save.
LCD Display Description
Changed
Return to previous page after displaying above message.
To cancel
▌To cancel, press I/O LOCAL Key or wait for 10 seconds.
Wait 10 sec
LCD Display Description
Cancel…
Read reference "1-6. Check after Power ON" "Earliest Setting Value"
Note