4-6. Calibration Using REMOTE INTERFACE
Explanation about calibration using Remote Interface.
Other commands cannot be used while Remote Calibration.
Connecting Measuring Instrument
▌Before calibration, connect like diagram 4-2.
▌Set up Interface in each instrument
▌Set-up Interface in each instrument.
▌Activate warming-up to all required instruments including power supply.
Remote Calibration Command Sequence
▌Read reference "6-6. Calibration Command" for SCPI Commands.
▌Please proceed with below sequence, otherwise there might be error occurs
which stops Remote Calibration.
▌If an error occurs Remote Calibration should be done from the first
▌If an error occurs, Remote Calibration should be done from the first.
Voltage CALIBRATION Operation
▌Switch OFF the Electronic Load.
▌Send Voltage Minimum Calibration command to power supply.
Sending Command
CAL:VOLT MIN
Sending Command
▌After 5 minutes, measure DVM Voltage.
▌Send measured voltage value to power supply voltage value.
For example, if measured value is 0.1234, send as below.
Sending Command
It is OK when digits are Over.
CAL:VOLT 0.1234
CAL:VOLT MIN
▌Send Voltage Maximum Calibration command to power supply.
Sending Command
▌After 5 minutes, measure DVM Voltage.
▌Send measured voltage value to power supply voltage value.
For example, if measure value is 9.123, send as below.
CAL:VOLT MAX
For example, if measure value is 9.123, send as below.
Sending Command
It is OK when digits are Over.
CAL:VOLT 9.123