3-5. Programming Over Current Protection(OCP)
If output current is higher than OCP Level, this function shut off the output current to protect
load from power supply. It is OCP Level Set-up & ON,OFF set-up below :
OCP Level Value below can be different.
▌Power Switch ON
Check "**OUTPUT OFF**" Mode is displayed after input Power ON.
▌Press PROTECT Key to Set-up the Level
LCD Display Description
PROTECT> 1.OVP
▌Use right cursor key to move on to OCP Menu.
LCD Display Description
PROTECT> 2.OCP
▌Press PROTECT Key once more in "PROTECT> 2 OCP" condition
▌Press PROTECT Key once more in PROTECT> 2.OCP condition.
LCD Display Description
OCP SET> 3.3000A
▌Use cursor key to change the value.
▌Use encoder switch to change Level value.
OCP Level cannot be set-up lower than present output current.
▌When it is done, Press PROTECT Key to save.
LCD Display Description
Changed
It returns when above message is displayed.
To cancel
▌To cancel, press I/O Local Key or wait for 10 seconds.
wait 10sec
LCD Display Description
Cancel…
wait 10sec
LCD Display Description
Cancel…
Read reference for earliest OCP set-up "1-6. Check after Power ON" "Earliest setting value".
If actual output current is higher than OCP Level 'OCP TRIPPED" will be displayed and all the
Note
If actual output current is higher than OCP Level, OCP TRIPPED will be displayed and all the
output voltage and currents will be shut-off. To print voltage and current again,
remove load and follow below sequence to clear the trip.
Trip can be occurred
First It can happen when OCP Level value is accidentally tripped.
Therefore, raise Level value to solve the problem.
Second There can be a trip when users used load which cause counter electromotive
Second
There can be a trip when users used load which cause counter electromotive
force such as Coil & Motors. Diagram 3-2 above shows how to prevent it.
- 23 -