FACT AUTO LOC?
FACT:AUTO-LOC?
It is command which checks Auto Key Lock setting value.
Return value
"0"
Disable Condition
"1"
Enable Condition
ex)fact:auto-loc?
return value "1"
FACT:OVP {DIS|ENA}
I i
hi h h
k
h h
OV f
i
i
bl
i
bl
It is command which checks whether OVP function is enabled or disabled.
> DIS
Disable, OVP function is disabled.
> ENA
Enable, OVP function is enabled.
ex) fact:ovp dis
Set-up OVP function to be disabled.
FACT:OVP?
It is command which checks whether OVP is enabled or disabled.
" "
Return value
"0"
Disable Condition
"1"
Enable Condition
ex)fact:ovp?
return value "0"
FACT:OCP {DIS|ENA}
It is command which checks whether OCP function is enabled or disabled.
> DIS
Disable, OCP function is disabled.
> ENA
Enable OCP function is enabled.
ex) fact:ocp dis
Set-up OCP Function to be disabled.
FACT:OCP?
It is command which checks whether OCP is enabled or disabled.
Return value
"0"
Disable Condition
"1"
Enable Condition
ex)fact:ocp?
return value "0"
FACT:ADC {5|20|50|100|300|1300}
It is command which set-up power supply voltage/current speed and accuracy.
> 5
Measuring voltage or current 5times/1sec, Highly-accurate
> 20
Measuring voltage or current 20times/1sec, Highly-accurate
> 50
Measuring voltage or current 50times/1sec, accurate
> 100
Measuring voltage or current1005times/1sec, accurate
> 300
Measuring voltage or current 300times/1sec, Measure hunting value
> 1300
Measuring voltage or current 1300times/1sec, Measure hunting value
ex) fact:ADC 5
Sampling speed is low, however, it is very accurate.
- 64 -