3-4. Programming Over Voltage Protection(OVP)
Function that protects by shutting off output voltage when the voltage is higher than OVP Level.
Read OVP Level Set-up Below.
Disable OVP function in Factory Mode.
OVP Level Value below can be different.
▌Power Switch ON
Check "**OUTPUT OFF**" Mode is displayed after input Power ON.
▌Press PROTECT Key to Set-up the Level
LCD Display Description
PROTECT> 1.OVP
▌Press PROTECT Key once more in "1.OVP" condition.
LCD Display Description
OVP SET> 9 900V
LCD Display Description
OVP SET> 9.900V
▌Use cursor key to locate the value.
▌Use Encoder switch to set-up the level.
OVP Level cannot be set less than outputting voltage.
▌Press PROTECT Key when it is done.
LCD Display Description
Changed
Returns after displaying above Message.
To cancel
▌To cancel, please press I/O LOCAL Key or wait for 10 seconds..
Wait 10 Sec
LCD Display Description
Cancel…
Note
Read reference about earliest OVP Set-up "1-6. Check after power ON" "Earliest Setting Value"
W
hen actual output voltage is higher than OVP Level value "
OVP TRIPPED
" Message
W
hen actual output voltage is higher than OVP Level value,
OVP TRIPPED
Message
will be displayed. It shut-off output voltage and current to prevent the sources.
To output voltage and current again, remove the load and follow the sequence below to
Clear the Trip. The Trip might occurred by :
First It can happen when OVP Level value is accidentally tripped.
Therefore, raise Level value to solve the problem.
Second There can be a trip when users used load which cause counter electromotive
force such as Coil & Motors. Diagram below shows how to prevent it.