5 10 CAL RESTORE
5-10. CAL-RESTORE
▌This is function which restores back-up calibration data to current system.
▌It can bring back calibration data when previous calibration data is deleted, or differed from
testing circumstances.
CAL-RESTORE Operation
▌Power Switch ON while pressing FACTORY Key.
▌Power Switch ON
▌
When
"
1. USER-MEM CLEAR
" Message is displayed, release FACTORY Key.
Holding
R l
▌
Use cursor key to move on to
"
8. CAL-RESTORE
" Menu.
▌Press FACTORY Key to Restore.
LCD Display Description
Done
Release
5-11. CAL-BACKUP
▌To obtain guarantee of high-accuracy, the product should be calibrated every 6 month at
institute. Also, to use without problem, it should be calibrated every 1 year. To prevent
worst happening, please back-up calibration data from the institute.
▌When back-up is activated, previous back-up description cannot be restored.
CAL-BACKUP Operation
▌Power Switch ON while pressing FACTORY Key.
▌Power Switch ON
Holding
▌
When
"
1. USER-MEM CLEAR
" Message is displayed, release FACTORY Key.
▌
Use cursor key to move on to
"
9. CAL-BACKUP
" Menu.
▌Press FACTORY Key to Back-up.
LCD Display Description
Done
Release
LCD Display Description
Done
- 53 -