▌If it is stabilized, calculate current value.
If used resist is 0.01Ω and measured voltage value is 0.0123mV,
the current value will be 0.00123A.
▌Use cursor key and encoder switch to input measured current to power supply.
LCD Display Description
A LOW 0 0012A
LCD Display Description
A-LOW 0.0012A
▌When it is done, press CAL Key to save.
LCD Display Description
ADC DATA 026CH
Readback value will be
displayed in HEX for 500ms
LCD Display Description
3.CAL-CURR LOW
Return to previous CAL page.
▌
Use cursor key to select
"
4.CAL-CURR HIGH
" and press CAL Key.
LCD Display Description
A-HIGH 3.2778A
Current can be different.
▌Wait for voltage shown on DVM Meter to be settled.
Wait for 5 minutes
▌If it is stabilized, calculate current value.
If used resist is 0.01Ω and measured voltage value is 32.791mV, current value will be
3.2791A.
▌Use cursor key and encoder switch to input measured current value.
A HIGH 3 2 91A
LCD Display Description
A-HIGH 3.2791A
▌When it is done, press CAL Key to save.
LCD Display Description
ADC DATA F500H
Readback value will be
displayed in HEX for 500ms
LCD Display Description
4.CAL-CURR HIGH
Return to previous CAL page.
▌Current Calibration is done and it will be applied from the next operation of
power supply.
- 42 -