Image Optimization
5-9
Operation
Adjust through the [iBeam] item in the soft menu or menu.
The system provides 3 levels of iBeam adjustment, off represents no
iBeam, and while 2 represents maximum iBeam optimization.
Effects
Images after iBeam processing can be optimized with less spot noise and
higher resolution, so that more details for the structure are revealed.
Impacts
When linear probe is being used, iBeam function is not available when
trapezoid function is turned on or B steer is adopted.
Image Merge
Description
In the Dual-split mode, when the images of the two windows use the same
probe type, depth, invert status, rotation status and magnification factor, the
system will merge the two images so as to extend the field of vision.
Operation
Turn on or off the function through the [Img Merge] item in the soft menu or
menu.
Impacts
iBeam is available only for linear probes.
The function is available in real-time imaging, freeze or cine review status.
Gray Map
Description
This function applies the gray correction to obtain the optimum images.
Operation
Select among the maps through the [Gray Map] item in the soft menu or
menu.
You can also adjust through the grayscale bar: move the cursor to the
grayscale bar and press <Set> on the control panel to adjust.
The system provides 8 groups of gray map.
Impacts
The function is available in real-time imaging, freeze or cine review status.
IP (Image Processing)
Description
IP is a combination of several image processing parameters, which is used
for a fast image optimization, the real-time group of which is displayed in
the image parameter area in the upper left corner of the screen.
The IP combination parameters include dynamic range, iClear, persistence
and smooth.
Operation
Select among the IP groups through the [IP] item in the soft menu or menu.
Or adjust it in the image parameter area.
The system provides 8 groups of IP combinations, and the specific value of
each parameter can be preset.
Colorize and Colorize Map
Description
Colorize function provides an imaging process based on color difference
rather than gray distinction.
Operation
Turn on or off the function through the [Colorize] item in the soft menu or
menu.
Select the colorize map through the [Colorize Map] item in the soft menu or
menu.
The system provides 10 colorize maps to be selected among.
Содержание DC-T6
Страница 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 10: ......
Страница 16: ......
Страница 28: ......
Страница 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Страница 178: ......
Страница 182: ......
Страница 236: ......
Страница 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Страница 300: ......
Страница 314: ......
Страница 320: ......
Страница 326: ......
Страница 330: ...C 4 Barcode Reader...
Страница 337: ...Barcode Reader C 11...
Страница 342: ......
Страница 347: ...P N 046 001523 01 V1 0...