5-16 Image Optimization
Operation
Rotate the <Color> knob clockwise to increase the gain, and anticlockwise
to decrease.
Or adjust it in the image parameter area.
The adjusting range is 0-100.
Impacts
Increasing the gain will increase the flow signal presented as well as noise,
while the signals may be missing when the gain is adjusted too low.
ROI Adjustment
Description
To adjust the width and position of ROI in Color mode.
Operation
When the ROI box is dotted line, roll the trackball to change its size.
When the ROI box is solid line, roll the trackball to change the position.
Press <Set> to switch between the solid line and the dotted line status.
You can set the default ROI position and size in "[System Preset]
→
[Image
Preset]
→
Color".
Impacts
The larger the ROI box, the lower the frame rate, and the lower the
resolution and color sensitivity.
Frequency
Description
Refers to the operating frequency in Color mode of the probe, the real-time
value of which is displayed in the image parameter area in the upper left
corner of the screen.
Operation
Select the frequency value through the [Frequency] item in the soft menu
or menu.
Or adjust it in the image parameter area.
Values of frequency vary by probes. Select the frequency value according
to the need of the detection depth and the current tissue characteristics.
Effects
The higher the frequency, the worse the axial resolution, and the better the
force of penetration.
Focus Position
Description
To adjust the color focus position against the ROI position.
Operation
Change the position through the [Focus Position] item in the soft menu or
menu.
B Display
Description
To turn on or off the B image display as Color imaging remains active.
Operation
Turn on or off the function through the [B Display] item in the soft menu or
menu.
Impacts
The function is available in real-time imaging, freeze or cine review status.
B/C Wide
Description
To set and constrain the maximum width of the B mode image to that of the
Color ROI.
Operation
Turn on or off the function through the [B/C Wide] item in the soft menu or
menu.
Содержание DC-T6
Страница 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 10: ......
Страница 16: ......
Страница 28: ......
Страница 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Страница 178: ......
Страница 182: ......
Страница 236: ......
Страница 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Страница 300: ......
Страница 314: ......
Страница 320: ......
Страница 326: ......
Страница 330: ...C 4 Barcode Reader...
Страница 337: ...Barcode Reader C 11...
Страница 342: ......
Страница 347: ...P N 046 001523 01 V1 0...