D-4 Electrical Safety Inspection
LIMITS
For BF
applied parts
z
100
μ
A in Normal Condition
z
500
μ
A in Single Fault Condition
D.8 Mains on Applied Part Leakage
The Mains on Applied Part test applies a test voltage, which is 110% of the mains voltage,
through a limiting resistance, to selected applied part terminals. Current measurements
are then taken between the selected applied part and earth. Measurements are taken
with the test voltage (110% of mains) to applied parts in the normal and reverse polarity
conditions.
The following outlet conditions apply when performing the Mains on Applied Part test.
z
Normal Polarity;
z
Reversed Polarity
LIMITS
z
For BF
applied parts: 5000
μ
A
D.9 Patient
Auxiliary
Current
Patient Auxiliary currents are measured between any selected Applied Part connector and
the remaining Applied Part connector s. All measurements may have a true RMS only
response.
The following outlet conditions apply when performing the Patient Auxiliary Current test.
z
normal polarity( Normal Condition);
z
reverse polarity( Normal Condition),
z
normal polarity with open neutral(Single Fault Condition);
z
reverse polarity with open neutral(Single Fault Condition).
z
normal polarity with open earth(Single Fault Condition);
z
reverse polarity with open earth(Single Fault Condition).
LIMITS
For BF
applied parts,
z
100
μ
A in Normal Condition
z
500
μ
A in Single Fault Condition
Note:
Make sure the safety analyzer is authorized comply with requirement of
IEC60601-1.
Follow the instructions of the analyzer manufacturer.
Содержание DC-T6
Страница 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 10: ......
Страница 16: ......
Страница 28: ......
Страница 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Страница 178: ......
Страница 182: ......
Страница 236: ......
Страница 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Страница 300: ......
Страница 314: ......
Страница 320: ......
Страница 326: ......
Страница 330: ...C 4 Barcode Reader...
Страница 337: ...Barcode Reader C 11...
Страница 342: ......
Страница 347: ...P N 046 001523 01 V1 0...