13-30 Probes and Biopsy
NGB-010 metal/needle detachable needle-guided bracket
(1) Put on the probe cover.
(2) Hold the probe by one hand, select proper needle-guided bracket, and hold it with
the other hand, and align the grooves of the needle-guided bracket with the tabs
of the probe, then push the needle-guided bracket forward, making the grooves of
the needle-guided bracket to match with the tabs of the probe. Set the
needle-guided bracket at the desired position, turn tightly the knob of fixing
needle-guided bracket to fix the needle-guided bracket.
(3) Check manually to confirm the needle-guided bracket is securely installed on the
probe.
(4) Select a proper guiding block and push it into the groove above the support of
needle-guided bracket, then turn tightly the knob of fixing the guiding block to fix
the guiding block on the support of needle-guided bracket.
(5) Insert a biopsy needle with the same specification as that of the guiding block into
the hole of the guiding block.
NGB-011
1. Inosculate the locating groove on the clamp with the two raised edges on the probe
head and align the locating pit of the clamp to the convex point on the probe head.
Содержание DC-T6
Страница 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 10: ......
Страница 16: ......
Страница 28: ......
Страница 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Страница 178: ......
Страница 182: ......
Страница 236: ......
Страница 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Страница 300: ......
Страница 314: ......
Страница 320: ......
Страница 326: ......
Страница 330: ...C 4 Barcode Reader...
Страница 337: ...Barcode Reader C 11...
Страница 342: ......
Страница 347: ...P N 046 001523 01 V1 0...