Probes and Biopsy
13-13
NOTE:
1.
Observe the graph here carefully to immerse the transducer. Only soak
parts of the transducer below the strain relief.
Compatible Disinfectants
Manufacturer
Trade Name
Procedures
Type
Pharmaceutical
Innovations,
Inc.
T-Spray II
Please refer to the instructions
provided by the manufacturer of the
solution for details.
Spray*
Parker
Laboratories
Inc.
PROTEX
TM
DISINFECTANT
SPRAY
Please refer to the instructions
provided by the manufacturer of the
solution for details.
Spray*
Metrex MetriZyme Please
refer to the instructions
provided by the manufacturer of the
solution for details.
Solution*
ASP
Cidex Activated
Dialdehyde
Solution
Please refer to the instructions
provided by the manufacturer of the
solution for details.
Solution
ASP Cidex
OPA
Please refer to the instructions
provided by the manufacturer of the
solution for details.
Solution
Nanosonics
Limited
TrophonSonex-HL
(Used with
Trophon EPR
Ultrasound Probe
Disinfector)
Please refer to the instructions
provided by the manufacturer of the
solution for details.
Solution**
* denotes if the disinfectant is not applicable for following probes: 4CD4, CW5s, P7-3,
CW2s
**
denotes if the disinfectant is not applicable for following probes: 4CD4, CW5s, P7-3,
CW2s
The probe C5-2 only uses the disinfectant PROTEX
TM
and CIDEX OPA.
Strain relief
Probe handle
Содержание DC-T6
Страница 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 10: ......
Страница 16: ......
Страница 28: ......
Страница 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Страница 178: ......
Страница 182: ......
Страница 236: ......
Страница 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Страница 300: ......
Страница 314: ......
Страница 320: ......
Страница 326: ......
Страница 330: ...C 4 Barcode Reader...
Страница 337: ...Barcode Reader C 11...
Страница 342: ......
Страница 347: ...P N 046 001523 01 V1 0...