5-64 Image Optimization
Angle
z
Function: to set the range for 4D imaging.
z
Adjusting range: 10°-70°, in increment of 5°.
Quality
z
Function: to adjust the image quality by changing the line density. Image quality
can affect the imaging speed that the better the image quality, the lower the
speed, so the frame rate is lower.
z
Range: Low2, Low1, Mid, High1, High2
Other parameters are the same as that of Smart 3D, for details; please refer to
“5.11.4.2 Smart 3D Acquisition Preparation”.
5.11.5.3 4D Image Acquisition
1. Optimize the 2D image of the desired region.
z
High contrast between the desired region and the surroundings.
z
Clear boundary of the desired region.
z
Little noise
2. Set ROI (Region of Interest) and curve VOI on the 2D image.
Roll the trackball to change the ROI size and position as well as curve VOI, press the
<Set> key to toggle among setting the ROI size, position or curve VOI.
Set ROI on the 2D image with the largest area of the target region. The imaging area
should be only a little larger than ROI; otherwise frame rate may be too low.
3. Click [Start] to begin the acquisition.
In 4D image acquisition, the probe performs swinging in the defined range of angle
automatically.
5.11.5.4 4D Real-time Display
In 4D real-time display status, the following operations can be performed (for the details,
please refer to the corresponding contents in Smart 3D.).
z
Set [Quality] and [Angle].
z
Set the image display format. The system can support three formats: Single, Dual
and Quad.
z
Set 3D image view direction by clicking [Direction].
z
Select section image to define the VOI.
z
Set the brightness, contrast and colorize map for 3D image.
z
Rotate and shift the 3D images, and reset the images to original status by clicking
[Reset].
z
View section image and 3D image.
z
Image rendering setting.
z
Save single frame image by a user-defined save key.
z
Other general operations like zooming, comment and body mark.
5.11.5.5 4D Cine Reviewing
In 4D real-time display mode, press <Freeze> on the control panel to enter the freeze
mode.
The following operations can be performed (for the details, please refer to the
corresponding contents in Smart 3D).
z
Set the image display format. The system can support three formats: Single, Dual
and Quad.
z
Select section image to define the VOI.
Содержание DC-T6
Страница 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 10: ......
Страница 16: ......
Страница 28: ......
Страница 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Страница 178: ......
Страница 182: ......
Страница 236: ......
Страница 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Страница 300: ......
Страница 314: ......
Страница 320: ......
Страница 326: ......
Страница 330: ...C 4 Barcode Reader...
Страница 337: ...Barcode Reader C 11...
Страница 342: ......
Страница 347: ...P N 046 001523 01 V1 0...