Image Optimization
5-17
Impacts
Frame rate will increase when the function is turned on.
Dual Live
Description
This function is used to display B image and Color image synchronously.
Operation
Click [Dual Live] in the soft menu or menu to turn on or off the function.
When the function is turned on, the window will be automatically switched
to the dual windows (one for B image, and the other for Color image).
Impacts
The function is available in real-time imaging, freeze or cine review status.
Steer
Description
The feature is used to adjust the ROI of color flow with different angles with
immobility of the linear probe.
Operation
Adjust through the [Steer] item in the soft menu or menu, or rotate the
multifunctional knob to adjust.
Values of steering angles vary by probes.
Effects
This function is used to adjust the scan angle of linear probes, so as to
change the angle between the transmitting beam and flow direction.
Impacts
Steer is available only for linear probes.
Line Density
Description
The function determines the quality and information of the image.
Operation
Adjust through the [Line Density] item in the soft menu or menu.
There are 2 levels of line density provided: UH, H, M, L.
Effects
The higher the line density, the higher the resolution.
Impacts
The higher the line density, the smaller the frame rate.
Packet Size
Description
This function is an indication of the ability to detect flow, which is used to
adjust the accuracy of color flow.
Operation
Adjust through the [Packet Size] item in the soft menu or menu.
There are 4 levels of packet size provided, 0 represents no packet size
control and the bigger the value the higher the sensitivity.
Effects
The higher the packet size, the more sensitive indication for low-velocity
flow.
Impacts
Increasing the packet size will lead to frame rate decrease.
Flow State
Description
This function is used for a fast image optimization by providing an
adjustment of 2 parameters, including scales and filters.
Operation
Adjust through the [Flow State] item in the soft menu or menu.
There are 3 levels provided to adjust: L, M, H.
Содержание DC-T6
Страница 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 10: ......
Страница 16: ......
Страница 28: ......
Страница 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Страница 178: ......
Страница 182: ......
Страница 236: ......
Страница 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Страница 300: ......
Страница 314: ......
Страница 320: ......
Страница 326: ......
Страница 330: ...C 4 Barcode Reader...
Страница 337: ...Barcode Reader C 11...
Страница 342: ......
Страница 347: ...P N 046 001523 01 V1 0...