5-20 Image Optimization
Priority
Description
This function is used to set levels of the flow display, to display the
grayscale signal or color signal.
Operation
Select the value through the [Priority] item in the soft menu or menu.
The adjusting range is 0-100%.
Effects
The higher the value, color signals are prior to be displayed; while the lower
the value, grayscale signals are prior to be displayed.
iTouch
Description
To optimize image parameters as per the current tissue characteristics for a
better image effect.
Operation
Press <iTouch> on the control panel to turn on the function, the symbol of
which will be displayed in the image parameter area of the screen.
5.6
Power Mode Image Optimization
Power mode provides a non-directionally display of blood flow in the form of intensity as
opposed to flow velocity.
DirPower (Directional Power Mode) provides the additional information of flow direction
towards or away from the probe.
5.6.1 Power Mode Exam Protocol
1. Select a high-quality image during B mode or B+ Color scanning, and adjust to place
the area of interest in the center of the image.
2. Press <Power> to enter B+Power mode. Use the trackball and <Set> to change
position and size of the Region of Interest (ROI).
3. Adjust the image parameters to obtain optimized images.
4. Perform other operations (e.g. measurement and calculation) if necessary.
Содержание DC-T6
Страница 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 10: ......
Страница 16: ......
Страница 28: ......
Страница 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Страница 178: ......
Страница 182: ......
Страница 236: ......
Страница 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Страница 300: ......
Страница 314: ......
Страница 320: ......
Страница 326: ......
Страница 330: ...C 4 Barcode Reader...
Страница 337: ...Barcode Reader C 11...
Страница 342: ......
Страница 347: ...P N 046 001523 01 V1 0...