Probes and Biopsy
13-29
d) Select a proper guiding block and push it into the groove above the angle
block, and clamp it tightly.
e) Screw the nut of the block to secure the block.
f) Insert a biopsy needle with the same specification as that of the guiding block
into the hole of the guiding block.
z
Plastic needle-guided bracket
a) Put on the sterile probe sheath.
b) Hold the probe by one hand, select the proper needle-guided bracket, and
hold it with the other hand. Align the narrow end tab of the needle-guided
bracket with the groove of the probe, then push the needle-guided bracket
forward, making the tabs and the grooves of the needle-guided bracket to
match with the grooves and tabs of the probe.
c) Check manually to confirm that the needle-guided bracket is securely
installed on the probe.
d) Select a proper guiding block and push it into the groove above the angle
block, and clamp it tightly.
e) Insert a biopsy needle with the same specification as that of the guiding block
into the hole of the guiding block.
Содержание DC-T6
Страница 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 10: ......
Страница 16: ......
Страница 28: ......
Страница 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Страница 178: ......
Страница 182: ......
Страница 236: ......
Страница 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Страница 300: ......
Страница 314: ......
Страница 320: ......
Страница 326: ......
Страница 330: ...C 4 Barcode Reader...
Страница 337: ...Barcode Reader C 11...
Страница 342: ......
Страница 347: ...P N 046 001523 01 V1 0...