ECG
8-3
Menu
Preset
Click [Menu Preset] in the Image Preset screen to enter the menu preset page, and
click “Physio” page tab to go for the setting.
Parameter
Description
Display
Function: to set if to display ECG trace on the screen.
Position
Function: to set the vertical position of the ECG trace on the image display.
Value: 0%-100%, in increments of 5%.
Display HR Function: to set if to display patient heart rate on the screen.
Gain
Function: to set the amplitude of ECG trace.
Value: 0-30, in increments of 1.
8.3 Parameter
Description
Parameter
Description
ECG
Function: to control the display of ECG trace in live status.
Value: On/Off, On is to display the trace, while Off is to hide the trace.
In scanning mode, turn off ECG display will stop ECG signal acquisition.
In freeze mode, this is used to show or hide the ECG trace.
ECG Gain
Function: to set the amplitude of ECG trace.
Method: click [Gain] in the soft menu or menu.
Value: 0-30, in increments of 1.
ECG
Position
Function: to set the vertical position of the ECG trace on the image display.
Method: click [ECG Pos] in the soft menu or menu.
Value: 0%-100%, in increments of 5%.
Speed
Function: change the speed of the physio trace.
Value: 1-4
The refresh speed of the physio signal on B image is independent from the
refresh speed of timeline refresh speed (M mode, PW and CW mode)
8.4 ECG
Review
When an image is frozen, the ECG waveform will be frozen at the same time. When
images are reviewed with the ECG electrodes connected, the ECG waveform is the
reference for time.
After the images are frozen, all real time images are in the status of linked review.
The frame position marker indicates the time relationship between the currently reviewed
2D image and the ECG trace or M/D image.
Содержание DC-T6
Страница 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 10: ......
Страница 16: ......
Страница 28: ......
Страница 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Страница 178: ......
Страница 182: ......
Страница 236: ......
Страница 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Страница 300: ......
Страница 314: ......
Страница 320: ......
Страница 326: ......
Страница 330: ...C 4 Barcode Reader...
Страница 337: ...Barcode Reader C 11...
Страница 342: ......
Страница 347: ...P N 046 001523 01 V1 0...