Image Optimization
5-69
Status
ROI Color
Tips
Appropriate Green None
Speed too high Red
Moving speed of the probe is too high.
Guidance and precautions for uniform motion:
z
Make sure that there is enough coupling gel along the scan path.
z
Always move the probe slowly and steadily.
z
Continuous contact is required throughout the length of the extended image.
Don’t lift the probe from the skin surface.
z
Always keep the probe perpendicular to the skin surface. Don’t rock, rotate or tilt
the probe during the scan.
z
The system accommodates a reasonable range of motion velocity. Don’t make
abrupt changes in speed of motion.
z
Deeper scans generally require reduced acquisition speed.
5.12.4 iScape Viewing
After completing the image acquisition, the system performs image splicing and then
enters the iScape viewing mode.
In iScape viewing mode, you can perform the following functions:
Image parameters setting, for details, please refer to “5.12.4.1 Image Parameters ”.
Image zooming, for details, please refer to “5.12.4.2 Image Zooming”.
Image rotation, for details, please refer to “5.12.4.3 Rotating the Image”.
Measurement, comment, and body mark, for details, please refer to “5.12.4.4
Measurement, Comment, and Body Mark”.
5.12.4.1 Image Parameters Setting
In image viewing mode, you can adjust the following parameters:
Image
size
Click [Actual Size] to display the image in its actual size.
Click [Fit Size] to display the image in an appropriate size according to the current
window. Also this is the system default format.
Colorize and colorize map
Click [Colorize] to activate the colorize function, and then change the effect by clicking
[Colorize Map] in the soft menu.
Ruler
Click [Ruler] to turn on or off the ruler display on the image.
5.12.4.2 Image Zooming
Press <Zoom> knob on the control panel to enter image zooming mode. Rotate the
<Zoom> knob to zoom in/out the panoramic image.
z
Rotate the knob clockwise to zoom in the image, the maximum factor is 4.
z
Rotate the knob anticlockwise to zoom out the image, the minimum factor is 0.01.
z
Roll the trackball to change position of the magnified image.
z
Press the <Zoom> knob again to exit the zoom mode.
z
When image displayed is bigger than the image area, the thumbnail is
automatically displayed.
Содержание DC-T6
Страница 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 10: ......
Страница 16: ......
Страница 28: ......
Страница 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Страница 178: ......
Страница 182: ......
Страница 236: ......
Страница 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Страница 300: ......
Страница 314: ......
Страница 320: ......
Страница 326: ......
Страница 330: ...C 4 Barcode Reader...
Страница 337: ...Barcode Reader C 11...
Страница 342: ......
Страница 347: ...P N 046 001523 01 V1 0...