Section 2
Introduction
25
MikroScan 7600PRO
Operator’s Manual
Catalog 11180-94
V15.4F 050406
[teMP]
Displays the temperature of the points
under the color bar. Does not, however,
allow emissivity settings to be changed or
displayed.
[teMP &
eMISS]
Displays the temperature and emissivity
of each point near the point cursor. Also
allows emissivity of each point to be set
individually..
[
∆
teMP]
Displays the specified point temperature
under the color bar. Also displays the
temperature difference
∆
t
between point
A
and point
b.
However, this mode does
not allow the emissivity to be corrected on
an individual basis.
[
∆
teMP &
eMISS]
Displays the specified point tempera-
ture under the color bar and displays the
emissivity setting of the point near the
point cursor. Also displays the temperature
difference
∆
t
between point
A
and point
b
. However, the emissivity is displayed
only when the emissivity of the point is
different from the emissivity of the whole
screen.
[bOx teMP]
Used to determine the Max, Min, and Avg.
temperatures within a specified area of a
thermal image. When creating a box on
a thermal image, the max, min, and avg.
temperatures within the area of the speci-
fied box are displayed.
[On]
The maximum, minimum, and average
temperature of the area defined by the box
is not displayed.
[OFF]
The maximum, minimum, and average
temperature of the area defined by the box
is displayed.
d)
[dISPLAy Menu)
Performs screen display settings for the
camera.
[nuC MOde]
Corrects uneven characteristics of the
Infrared detector elements (UFPA). Note:
the main unit internally has a reference
blackbody source, which has an even tem-
perature surface. NUC operation corrects
the unevenness of characteristics for all
elements of UFPA by the reference black-
body source. Also an internal accurate
thermometer monitors the temperature of
the reference blackbody source and cali-
brates the unit during the NUC Operation.
Содержание 7600PRO
Страница 2: ......