Section 4
Basic Operation
98
MikroScan 7600PRO
Operator’s Manual
Catalog 11180-94
V15.4F 050406
4.4 Switching Between Ranges
When the
L/S LInK
mode s selected whle n
RAnGe 1
and the temperature level
exceeds the upper lmt of
RAnGe 1
, the instrument will automatically jump to
RAnGe 2
. When the
L/S LInK
mode s selected whle n
RAnGe 2
and the temperature
level exceeds the lower lmt of
RAnGe 2
, the instrument will automatically jump to
RAnGe 1
.
4.4.1 Using the Manual Range Formatting Option with L/S Link
1) Follow steps 1-4 for Selecting the Temperature Range Mode
Setting (See Section 4.3.2 of this manual).
In step 3 of Selecting the Temperature Range Mode Setting, be
sure to select
[L/S LInK]
.
[L/S LInK RAnGe MOde SeLeCted]
2) While in
RANGE 1
, Toggle the
jOyStICK
up until the temperature
level exceeds the upper limit of
RANGE 1
.
[RAnGe PROCeSSInG MeSSAGe]
NOTE
The Maximum Tempera-
ture Level for Range 1
is 120°C (248°F), while
the Minimum Tempera-
ture Level for Range 1 is
-40°C (-40°F).
The Maximum Tempera-
ture Level for Range 2 is
500°C (932°F), while the
Minimum Temperature
Level for Range 2 is 0°
(32°F).
NOTE
The Maximum Tempera-
ture Level for Range 1
is 120°C (248°F), while
the Minimum Tempera-
ture Level for Range 1 is
-40°C (-40°F).
The Maximum Tempera-
ture Level for Range 2 is
500°C (932°F), while the
Minimum Temperature
Level for Range 2 is 0°
(32°F).