Section 6
Data Analysis
200
MikroScan 7600PRO
Operator’s Manual
Catalog 11180-94
V15.4F 050406
4) Press the
C
[CAnCeL/VISIbLe]
button to delete the selected cursor
point.
POInt LAbeL
POInt eMISSIVIty VALue
P. teMP &
eMISS
POSItIOn [a]
P. teMP & eMISS
POSItIOn [b]
When a selected cursor point has been deleted, the remaining
cursor ponts are relabeled.
5) Follow the previous steps 3-4 to delete additional cursor points.
POInt LAbeL
POInt eMISSIVIty VALue
P. teMP & eMISS
POSItIOn [a]
6) Press the
jOyStICK [e]
button to set the remaining cursor point
location(s) and return to the main display.
Pressng
jOyStICK [e]
button wll lock the new cursor ponts
to ther current postons.
NOTE
When an emissivity value
has been set for a cursor
point, that emissivity value
stays with the designated
point until it is changed.
Deleting Emissivity val-
ues will have no affect on
new cursor points. In other
words, when a new cursor
point is created, make sure
that it has not taken on an
old emissivity value.
Emissivity Values of points
do not automatically default
back to E1.00.