Section 6
Data Analysis
197
MikroScan 7600PRO
Operator’s Manual
Catalog 11180-94
V15.4F 050406
6.5.4.2 Setting Emissivity Values for One or More Points
1) Follow steps 1-5 for Setting the Display to Show Emissivity Effects
on Various Points (See Section 6.5.4.1).
2) Toggle the
jOyStICK
up or down as needed to scroll through the
highlighted cursor point labels to select the cursor point that needs
modification.
POInt LAbeL HIGHLIGHted
POInt eMISSIVIty VALue
eMISSIVIty FOR
P.teMP POSItIOn
[c]
3) Press the
S
[SAVe]
button to highlight the
[eMISSIVIty VALue]
of the cursor point.
POInt LAbeL
POInt eMISSIVIty VALue
HIGHLIGHted
eMISSIVIty
FOR P.teMP
POSItIOn [c]
4) Toggle the
jOyStICK
up or down as needed to increase or decrease
the emissivity value of the selected cursor point.
POInt LAbeL
POInt eMISSIVIty VALue
HIGHLIGHted
P.teMP
POSItIOn [c]
NOTE
[TEMP & EMISS]
Displays the temperature and
emissivity of each point near
the point cursor. Also allows
emissivity of each point to be
set individually.
[
TEMP & EMISS]
Displays the specified point
temperature under the color bar
and displays the emissivity set-
ting of the point near the point
cursor. Also displays the tem-
perature difference
t
between
point
A
and point
b
. However,
the emissivity is displayed only
when the emissivity of the point
is different from the emissivity
of the whole screen.
NOTE
When an emissivity value
has been set for a cursor
point, that emissivity value
stays with the designated
point until it is changed.
Deleting Emissivity val-
ues will have no affect on
new cursor points. In other
words, when a new cursor
point is created, make sure
that it has not taken on an
old emissivity value.
Emissivity Values of points
do not automatically default
back to E1.00.