5.1.2
Observation Period, Measuring Time, End at Over-range,
Specifying Power, and Class Options
Menu
Use this menu for selecting a device operation type to set the observation period
(measuring time). The content is used for the report printout. It does not directly
affect measurements.
Device operation types
Use this menu for setting the measuring time. The setting range is 1 to 9600
seconds. Set the value, in view of the operation cycle of the EUT and the device
operation type that is selected using the F1 key.
If you set the standard to IEC Ed5.0/Ed2.1 or IEC Ed5.0/Ed1.0, select Class C, and
set the limit values to 3rd/5th/Current Wave, the measurement time will be 0.2
seconds. The Measurement Time setting will be disabled.
If the measured value exceeds the set voltage range or current range (overrange), an
error message will be displayed and the measurement will be terminated. However,
the data up to the point when the overrange occurs can be analyzed. The judgment
will be FAIL.
Use this menu to specify how to handle the power value of the EUT. Select the
measured or specified value in the sub-menu. Enter a numeric value in the power
specification value (W).
F1
Observation
Period
(Tobs)
Quasi-
Stationary
Short
Cycle
Random
Long
Cycle
F 1
F 1
F 2
F 2
F 3
F 3
F 4
F 4
F 5
F 5
F 6
F 6
Page
2/3
See
Quasi-
Stationary
Tobs (observation period) to be set has a continuation period long enough to
satisfy a repeatability request to device operation type “Quasi-stationary.”
Short Cycle
Tobs (observation period) to be set has 10 or more fundamental cycles or
a continuation period long enough to satisfy a repeatability request to
device operation type “Short cycle.” It is very close to an integer multiple
of the device operation cycle.
Random
Tobs (observation period) to be set has a continuation period long enough
to satisfy a repeatability request to device operation type “Random.”
Long Cycle
Tobs (observation period) to be set is the total program cycle of a device to
device operation type “Long cycle” or 2.5 minutes that is typically assumed
as the operation period in which the maximum THC is generated.
F2
Measurement
Time (s)
Numeric
value
F3
Overrange
Abort
Enabled
See
F4
Definition of
Power
Sub Menu
Definition of
Power
Specified
Power (W)
Содержание KHA1000
Страница 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Страница 8: ...vi KHA1000 ...
Страница 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 22: ...xx KHA1000 ...
Страница 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Страница 30: ...1 8 KHA1000 ...
Страница 68: ...3 20 KHA1000 ...
Страница 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Страница 156: ...5 54 KHA1000 ...
Страница 183: ...Other Measurements This chapter explains measurements for other than harmonic current and voltage fluctuation tests 7 ...
Страница 193: ...Remote Control This chapter explains how to connect the remote interface and how to control it using SCPI commands 8 ...
Страница 260: ...9 6 KHA1000 ...
Страница 261: ...Specifications This chapter explains the specifications of this product 10 ...
Страница 272: ...I 4 KHA1000 ...
Страница 273: ......