8-34
KHA1000
VF:DMAX
Sets the limit value of dmax (maximum relative voltage fluctuation) in d
measurement (voltage fluctuation test).
Command
VF:DMAXlimit {4|6|7}
VF:DMAXlimit?
Parameter
Value:
4, 6, or 7 (The default is 6.)
Unit:
PCT
When *RST is sent, settings are made as shown in Table 8-4.
Response
Returns the limit value of dmax (maximum relative voltage fluctuation) in d
measurement (voltage fluctuation test) in the <NR3> format.
VF:DCO
Sets the d measurement count.
Command
VF:DCOunt <NRf>
VF:DCOunt?
Parameter
Value:
3 to 24 (The default is 24.)
Unit:
None
When *RST is sent, settings are made as shown in Table 8-4.
Response
Returns the d measurement count in the <NR3> format.
VF:DTIM
Sets the d measurement time.
Command
VF:DTIMe{<numeric>|MIN|MAX}
VF:DTIMe?
Parameter
Value:
30 to 180 (The default is 60.)
Unit:
S
When *RST is sent, settings are made as shown in Table 8-4.
Response
Returns the d measurement time in the <NR3> format.
Содержание KHA1000
Страница 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Страница 8: ...vi KHA1000 ...
Страница 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 22: ...xx KHA1000 ...
Страница 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Страница 30: ...1 8 KHA1000 ...
Страница 68: ...3 20 KHA1000 ...
Страница 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Страница 156: ...5 54 KHA1000 ...
Страница 183: ...Other Measurements This chapter explains measurements for other than harmonic current and voltage fluctuation tests 7 ...
Страница 193: ...Remote Control This chapter explains how to connect the remote interface and how to control it using SCPI commands 8 ...
Страница 260: ...9 6 KHA1000 ...
Страница 261: ...Specifications This chapter explains the specifications of this product 10 ...
Страница 272: ...I 4 KHA1000 ...
Страница 273: ......