4-6
KHA1000
4.2.2
Collective Power-On/Off
The power supply of the test system is collectively turned on/off using an external
power switch. The customer is requested to prepare the external power switch. Use
the external power switch with the following or larger rated current.
Table 4-2
Rated current of external power switch
Power-on of the test system
1.
Turn on the external power switch.
The power supply for this product, the AC Power Supply, and the Line Impedance
Network are turned on.
This product communicates with the AC Power Supply. If communication fails,
“Disabled” is automatically assumed as the system setting for “
AC Pow. Control
” of
this product.
Power-off of the test system
1.
Press the EXT CONT key.
The EXT CONT display appears.
2.
Select OFF in “F1 Key (AC Pow. OUTPUT)” from the menu.
The AC Power Supply output is turned off. The icon in the upper part of the display
shows a turned-off light bulb.
3.
Turn off the external power switch.
The power supply for this product, the AC Power Supply, and the Line Impedance
Network are turned off.
External
power switch
Condition of AC
input voltage
AC power supply
*1
*1.
For other PCR-LA, PCR-LE AC Power Supply, contact your Kikusui distributor or agent.
PCR1000LE
PCR1000LA
PCR2000LE
PCR2000LA
PCR4000LE
PCR4000LA
Rated current
90 V to 132 V
30 A
50 A
100 A
170 V to 250 V
15 A
30 A
50 A
Содержание KHA1000
Страница 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Страница 8: ...vi KHA1000 ...
Страница 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 22: ...xx KHA1000 ...
Страница 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Страница 30: ...1 8 KHA1000 ...
Страница 68: ...3 20 KHA1000 ...
Страница 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Страница 156: ...5 54 KHA1000 ...
Страница 183: ...Other Measurements This chapter explains measurements for other than harmonic current and voltage fluctuation tests 7 ...
Страница 193: ...Remote Control This chapter explains how to connect the remote interface and how to control it using SCPI commands 8 ...
Страница 260: ...9 6 KHA1000 ...
Страница 261: ...Specifications This chapter explains the specifications of this product 10 ...
Страница 272: ...I 4 KHA1000 ...
Страница 273: ......