KHA1000
2-13
2
In
sta
lla
tio
n
a
n
d
P
re
p
ara
tio
n
fo
r U
se
2.5
Connecting the Test System
Use the Line Impedance Network in harmonic current tests (JIS C61000-3-2(2019),
JIS C61000-3-2(2005) and JIS C610003-2(2003)) and a voltage fluctuation test (IEC
61000-3-3 (Ed2.0)).
The impedance is not used in IEC 61000-3-2 (Ed5.0), IEC 61000-3-2 (Ed3.0), and
IEC 61000-3-2 (Ed2.2). In JIS C61000-3-2(2003), the impedance is always used. In
JIS C61000-3-2(2005) and JIS C61000-3-2(2019), the impedance may be used if
test results have variations (optional). In a voltage fluctuation test (IEC 61000-3-3
(Ed2.0)), the impedance is used to measure current fluctuations that are generated
by voltage fluctuations and flicker.
•
To prevent an electric shock, be sure to remove the power cord of a
devices used in a test system or turn off the POWER switch.
Setting line impedance values
•
PCR-WE cannot be combined with a standard line impedance network. For the
line impedance network for PCR-WE, contact your Kikusui distributor or agent.
WARNING
Standard
Setting Line Impedance Network
IEC 61000-3-2(Ed5.0)
IEC 61000-3-2(Ed3.0)
IEC 61000-3-2(Ed2.2)
Bypass (THRU)
*1
*1.
Bypass (THRU) for the LIN1020JF Line Impedance Network
OUT (THRU) for the LIN40MA-PCR-L
JIS C61000-3-2(2005)
JIS C61000-3-2(2019)
Bypass (THRU)
*1
,
or
Z1(0.4
+ 0.37 mH): Nominal voltage 100 V (single phase)
Z2(0.38
+ 0.46 mH): Nominal voltage 200 V (single phase)
JIS C61000-3-2(2003)
Z1(0.4
+ 0.37 mH): Nominal voltage 100 V (single phase)
Z2(0.38
+ 0.46 mH): Nominal voltage 200 V (single phase)
IEC 61000-3-3(Ed2.0)
0.4
+ j0.25
Содержание KHA1000
Страница 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Страница 8: ...vi KHA1000 ...
Страница 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 22: ...xx KHA1000 ...
Страница 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Страница 30: ...1 8 KHA1000 ...
Страница 68: ...3 20 KHA1000 ...
Страница 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Страница 156: ...5 54 KHA1000 ...
Страница 183: ...Other Measurements This chapter explains measurements for other than harmonic current and voltage fluctuation tests 7 ...
Страница 193: ...Remote Control This chapter explains how to connect the remote interface and how to control it using SCPI commands 8 ...
Страница 260: ...9 6 KHA1000 ...
Страница 261: ...Specifications This chapter explains the specifications of this product 10 ...
Страница 272: ...I 4 KHA1000 ...
Страница 273: ......