KHA1000
8-47
8
R
em
o
te
C
o
n
tr
o
l
TRIG:SEQ2:LEV
TRIG:RUSH:LEV
Each sets the current trigger level for in-rush current measurement.
Command
TRIGger:SEQuence2:LEVel {<numeric>|MIN|MAX}
TRIGger:SEQuence2:LEVel?
TRIGger:RUSH:LEVel {<numeric>|MIN|MAX}
TRIGger:RUSH:LEVel?
Parameter
Value:
0.1 to 80 (The default is 0.1.)
Unit:
A
When *RST is sent, settings are made as shown in Table 8-4.
Response
Returns the current trigger level in the <NR3> format.
TRIG:SEQ3:SOUR
TRIG:TEST:SOUR
Each sets the conditions for actually starting measurement for test operation after
reception of the INIT:SEQ3 command.
Command
TRIGger:SEQuence3:SOURce {BUS|INTernal}
TRIGger:SEQuence3:SOURce?
TRIGger:TEST:SOURce {BUS|INTernal}
TRIGger:TEST:SOURce?
Parameter
Value:
BUS
Start measurement after waiting for a software trigger
(*TRG, IEEE488.1 get (Group Execute Trigger)).
IMMediate
Start measurement immediately (Default)
When *RST is sent, settings are made as shown in Table 8-4.
Response
Returns the current setting of the trigger source.
To start the second or a subsequent test in MANual mode set by the VF:METH
command (d measurement, voltage fluctuation test), use a software trigger (*TRG,
IEEE 488.1 get) regardless of the setting of TRIG:SEQ3:SOUR (TRIG:TEST:SOUR).
Содержание KHA1000
Страница 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Страница 8: ...vi KHA1000 ...
Страница 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 22: ...xx KHA1000 ...
Страница 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Страница 30: ...1 8 KHA1000 ...
Страница 68: ...3 20 KHA1000 ...
Страница 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Страница 156: ...5 54 KHA1000 ...
Страница 183: ...Other Measurements This chapter explains measurements for other than harmonic current and voltage fluctuation tests 7 ...
Страница 193: ...Remote Control This chapter explains how to connect the remote interface and how to control it using SCPI commands 8 ...
Страница 260: ...9 6 KHA1000 ...
Страница 261: ...Specifications This chapter explains the specifications of this product 10 ...
Страница 272: ...I 4 KHA1000 ...
Страница 273: ......