KHA1000
10-3
10
S
p
ec
ific
atio
n
s
Current measurement function
Power measurement function
Frequency measurement function
Item
Specifications
Range
0.5 A/1 A/2 A/5 A/10 A/20 A
Allowable crest
factor
0.5 A – 10 A
range
4 (0.5 A – 10 A range)
20 A range
*1
*1.
In-rush current measurement is excluded.
2.5 (20 A range)
Display item
*2
*2.
Accuracy is not specified for ± Peak.
TrueRMS and ± peak
Accuracy
*3
*3.
Accuracy is not specified for ± Peak. PLL is in the Lock state, in the effective input range, and with
anti-alias filter cutoff frequency 6 kHz.
45 Hz
–
65 Hz
0.5 A
range
± (0.5 % of rdng + 0.2 % of
range)
1 A – 20
A range
± (0.5 % of rdng + 0.1 % of
range)
Resolution
(RMS/Peak)
0.5 A range
0.0001 A/ 0.0001 A
1 A range
0.0001 A/ 0.0002 A
2 A range
0.0001 A/ 0.0004 A
5 A range
0.0001 A/ 0.001 A
10 A range
0.001 A/ 0.002 A
20 A range
0.001 A/ 0.004 A
Effective input
range
*4
*4.
Accuracy is not specified for the outside of range.
0.5 A range
50 % to 110 % of rated range
1 A – 20 A range
20 % to 110 % of rated range
Display of
excess input
0.5 A – 5 A range “Over” is displayed when the peak value exceeds the display
range or the RMS value exceeds 6.5000 Arms.
10 A,20 A range
“Over” is displayed when the peak value exceeds the display
range or the RMS value exceeds 24.050 Arms.
Item
Specifications
Display item
*1
*1.
The range is indicated by the maximum effective input voltage or maximum effective input current.
Real power (W) ---
Apparent power
(VA)
*2
*2.
Calculated from voltage, current, and active power. Accuracy follows that of voltage, current, or real
power.
VA = V × A
Reactive power
(var)
*1
var =
Power factor
(PF)
*1
PF = W/(VA)
Accuracy of
active power
*3
*3.
Frequency: 45 Hz to 65 Hz, voltage and current: sine wave, in the effective input range, power
factor 1.00.
P
150 W
± 1 % of
range
P < 150 W
± 1.5 W
Item
Specifications
Measurement range
*1
*1.
Applicable only in PLL lock mode.
45 Hz – 65 Hz
Accuracy
*1
± (0.15 % of rdng + 2 digit)
(VA)
2
−W
2
Содержание KHA1000
Страница 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Страница 8: ...vi KHA1000 ...
Страница 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 22: ...xx KHA1000 ...
Страница 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Страница 30: ...1 8 KHA1000 ...
Страница 68: ...3 20 KHA1000 ...
Страница 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Страница 156: ...5 54 KHA1000 ...
Страница 183: ...Other Measurements This chapter explains measurements for other than harmonic current and voltage fluctuation tests 7 ...
Страница 193: ...Remote Control This chapter explains how to connect the remote interface and how to control it using SCPI commands 8 ...
Страница 260: ...9 6 KHA1000 ...
Страница 261: ...Specifications This chapter explains the specifications of this product 10 ...
Страница 272: ...I 4 KHA1000 ...
Страница 273: ......