KHA1000
6-23
6
V
o
lta
g
e C
h
an
g
es
a
n
d
F
lu
ctu
atio
n
s a
n
d
F
lic
ke
r T
es
t
Fig.6-8
Example of report printout (measurement of voltage
changes caused by manual switching)
Company
Model name
PXZ1600
Type
Prototype
Serial No.
12345678
Operating mode
Date of test
2021/4/16 18:54:14
Climatic condition
Memo
Example
Power analyzer
Supply Source
Reference Impedance
Test Data of Voltage Change Caused by Manual Switching
FINAL TEST RESULT
PASS *1*2
Nominal Voltage
100 V
Nominal Frequency
50 Hz
Measuring Time
90 s
Voltage Change Caused by Manual Switching TEST REPORT
Test Engineer
Type of test
d Measurement Margin 10 %
dmax(%)
dmax(%)
Limit
Average
dc(%)
Tmax(ms)
dc(%)
Tmax(ms)
Limit
Maximum
Seg. 1
0.440
0.240
0
Seg. 2
0.440
0.070
0
Seg. 3
0.580
0.200
0
Seg. 4
-.---
-.---
------
Seg. 5
-.---
-.---
------
Seg. 6
-.---
-.---
------
Seg. 7
-.---
-.---
------
Seg. 8
-.---
-.---
------
Seg. 9
-.---
-.---
------
Seg.10
-.---
-.---
------
Seg.11
-.---
-.---
------
Seg.12
-.---
-.---
------
Seg.13
-.---
-.---
------
Seg.14
-.---
-.---
------
Seg.15
-.---
-.---
------
Seg.16
-.---
-.---
------
Seg.17
-.---
-.---
------
Seg.18
-.---
-.---
------
Seg.19
-.---
-.---
------
Seg.20
-.---
-.---
------
Seg.21
-.---
-.---
------
Seg.22
-.---
-.---
------
Seg.23
-.---
-.---
------
Seg.24
-.---
-.---
------
4.000
0.440
3.300
0.240
500
0
*1 Limits apply for the result at the condition of nominal voltage 230V and nominal frequency 50Hz.
5
*2 The values for the final test result are more than Margin value and less than or equal to Limit value.
EN 61000-3-3(2013)/A1(2019)
KHA1000,Ver2.00
*IEC 61000-3-3:Ed3.1(2017)
EN 61000-4-15(2011)
*IEC 61000-4-15:Ed2.0(2010)
S
d
d
T
T
T
1
2
4
3
6
8
9
10
ASS *1*2
P
14
15
16
)
T
9)
Содержание KHA1000
Страница 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Страница 8: ...vi KHA1000 ...
Страница 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 22: ...xx KHA1000 ...
Страница 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Страница 30: ...1 8 KHA1000 ...
Страница 68: ...3 20 KHA1000 ...
Страница 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Страница 156: ...5 54 KHA1000 ...
Страница 183: ...Other Measurements This chapter explains measurements for other than harmonic current and voltage fluctuation tests 7 ...
Страница 193: ...Remote Control This chapter explains how to connect the remote interface and how to control it using SCPI commands 8 ...
Страница 260: ...9 6 KHA1000 ...
Страница 261: ...Specifications This chapter explains the specifications of this product 10 ...
Страница 272: ...I 4 KHA1000 ...
Страница 273: ......