Common specifications of input section
Voltage measurement function
Item
Specifications
Maximum input
voltage
Continuous
300 Vrms/560 Vpeak (CAT I), 250 Vrms (CAT II)
Maximum input
current
Continuous
24 Arms/50 Apeak
Within 20 ms
80 Apeak
Voltage input impedance
About 4.7 M
Instrument loss
(between SOURCE and LOAD)
6 m
or less
Coupling mode
HA/Vf mode
Fixed to DC coupling
OTHER mode
DC or AC coupling (optional)
*1
*1.
The accuracy of this specification is not specified for AC coupling.
Item
Specifications
Range
*1
150 V/300 V
Allowable crest factor
*1
*1.
Up to ± 560 Vpeak
2
Display item
*2
*2.
Accuracy is not specified for
±
Peak.
TrueRMS and ± peak
Accuracy
*3
*3.
Accuracy is not specified for ± Peak. PLL is in the Lock state, in the effective input range, and with
anti-alias filter cutoff frequency 6 kHz.
± (0.4 % of rdng+0.04 % of
range)
Resolution
150 V range
0.01 V
300 V range
0.02 V
Effective input range
*4
*4.
Accuracy is not specified for outside of range.
60 % to 110 % of the range
Excess input display
“Over” is displayed when the peak value exceeds the display
range.
Fundamental wave frequency
range
45 Hz to 65 Hz
Содержание KHA1000
Страница 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Страница 8: ...vi KHA1000 ...
Страница 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 22: ...xx KHA1000 ...
Страница 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Страница 30: ...1 8 KHA1000 ...
Страница 68: ...3 20 KHA1000 ...
Страница 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Страница 156: ...5 54 KHA1000 ...
Страница 183: ...Other Measurements This chapter explains measurements for other than harmonic current and voltage fluctuation tests 7 ...
Страница 193: ...Remote Control This chapter explains how to connect the remote interface and how to control it using SCPI commands 8 ...
Страница 260: ...9 6 KHA1000 ...
Страница 261: ...Specifications This chapter explains the specifications of this product 10 ...
Страница 272: ...I 4 KHA1000 ...
Страница 273: ......