8-36
KHA1000
8.7.4
Measurement Functions
<meter_fn>
This product has 15 measurement functions for measuring current, harmonic
current, flicker, frequency, power, power factor, and voltage.
In this manual, a third-level node of FETC?, READ?, or MEAS? may be written as
<meter_fn>. Refer to Table 8-5 and replace <meter_fn> with the necessary
measurement method and code it.
Table 8-5
Specified value of <meter_fn>
Symbols used in Enable mode
•
HA:
Harmonic measurement mode
•
VF:
Voltage fluctuation measurement mode
•
BASIC: Basic measurement
•
FFT:
FFT analyzer
•
RUSH: In-rush current measurement
The response data is generated after waiting until the measurement is complete for
the FETC: <meter_fn>?, READ: <meter_fn>?, or MEAS: <meter_fn>? query
immediately after transmission of the INIT command or READ: <meter_fn>? /
MEAS: <meter_fn>? query.
<meter_fn.>
Measured value
Response
unit
Enable/disable mode
( : Enable, : Disable)
HA
*1
VF
*1
BASIC
FFT
RUSH
CURRent:AC
Alternate current
A
CURRent:AMPLitude:MAXimum
Maximum current
A
CURRent:AMPLitude:MINimum
Minimum current
A
CURRent:HARMonic[:AMPLitude] Harmonic current
A
CURRent:HARMonic:POHC
POHC
A
CURRent:HARMonic:THC
THC
A
FLICker:ST
Momentary flicker
–
FREQency
Frequency
Hz
POWer:AC[:REAL]
Real power
W
POWer:AC:APParent
Apparent power
VA
POWer:AC:REACtive
Reactive power
var
POWer:AC:PFACtor
Power factor
–
VOLTage:AC
Alternate current
voltage
V
VOLTage:AMPLitude:MAXimum
Maximum voltage
V
VOLTage:AMPLitude:MINimum
Minimum voltage
V
*1.
Disabled during testing or analysis in the test status (setting, test, or analysis).
Содержание KHA1000
Страница 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Страница 8: ...vi KHA1000 ...
Страница 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 22: ...xx KHA1000 ...
Страница 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Страница 30: ...1 8 KHA1000 ...
Страница 68: ...3 20 KHA1000 ...
Страница 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Страница 156: ...5 54 KHA1000 ...
Страница 183: ...Other Measurements This chapter explains measurements for other than harmonic current and voltage fluctuation tests 7 ...
Страница 193: ...Remote Control This chapter explains how to connect the remote interface and how to control it using SCPI commands 8 ...
Страница 260: ...9 6 KHA1000 ...
Страница 261: ...Specifications This chapter explains the specifications of this product 10 ...
Страница 272: ...I 4 KHA1000 ...
Страница 273: ......