KHA1000
8-33
8
R
em
o
te
C
o
n
tr
o
l
Parameter
Value:
100, 120, 200, or 230 (The default is 230.)
Unit:
V
When *RST is sent, settings are made as shown in Table 8-4.
When HARM:STAN “{IEC:ED22:ED20 | IEC:ED22:ED10 | IEC:ED30:ED20 |
IEC:ED30:ED10
| IEC:ED30A2:ED20A1 | IEC:ED30A2:ED10
}” is sent, the nominal voltage
is set to 230.
Response
The nominal voltage value is returned in the <NR3> format.
VF:METH
Sets the d measurement method.
Command
VF:METHod {AUTO|MANual}
VF:METHod?
Parameter
Value:
AUTO
Pst Auto (Default)
MANual
Manual
When *RST is sent, settings are made as shown in Table 8-4.
Response
Returns the d measurement method.
VF:DMAR
Sets the margin for the limit value of dc, dmax, or Tmax (d(t) > 3.3 %). Set a relative
value of the reference limit value (100). The setting range is 10 to 100. For instance,
specify 80 to set the margin to 80 % of the standard limit value. For Tmax (d(t) > 3.3
%), set a margin for the limit value of the time exceeding Tmax (d(t) > 3.3 %).
Command
VF:DMARgin {<numeric>|MIN|MAX}
VF:DMARgin?
Parameter
Value:
10 to 100 (The default is 100.)
Unit:
PCT
When *RST is sent, settings are made as shown in Table 8-4.
Response
Returns the margin for the limit value of dc, dmax, or Tmax (d(t) > 3.3 %) in the <NR3>
format.
Содержание KHA1000
Страница 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Страница 8: ...vi KHA1000 ...
Страница 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 22: ...xx KHA1000 ...
Страница 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Страница 30: ...1 8 KHA1000 ...
Страница 68: ...3 20 KHA1000 ...
Страница 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Страница 156: ...5 54 KHA1000 ...
Страница 183: ...Other Measurements This chapter explains measurements for other than harmonic current and voltage fluctuation tests 7 ...
Страница 193: ...Remote Control This chapter explains how to connect the remote interface and how to control it using SCPI commands 8 ...
Страница 260: ...9 6 KHA1000 ...
Страница 261: ...Specifications This chapter explains the specifications of this product 10 ...
Страница 272: ...I 4 KHA1000 ...
Страница 273: ......