8-60
KHA1000
8.9
Error Messages
Command error list
An error in the [-199, -100] range indicates that an IEEE 488.2 syntax error was
detected by the syntax parser of the measuring instrument. If an error in this class
occurs, the command error bit (bit 5) of the event status register is set.
Table 8-11 Command error list
Error code
Explanation of error message
-100 Command error
Generic syntax error.
-101 Invalid character
A syntactic element contains a character which is invalid.
-102 Syntax error
An unrecognized command or data type was encountered.
-103 Invalid separator
The parser was expecting a separator and encountered an illegal
character.
-104 Data type error
The parser recognized a data element different than one allowed.
-105 GET not allowed
A Group Execute Trigger was received within a program message.
-108 Parameter not allowed
More parameters were received than expected for the header.
-109 Missing parameter
Fewer parameters were received than required for the header.
-110 Command header error
An error was detected in the header.
-120 Numeric data error
Generated when parsing a data element which appears to be
numeric, including the nondecimal numeric types.
-130 Suffix error
Generated when parsing a suffix.
-131 Invalid suffix
The suffix does not follow the syntax, or the suffix is inappropriate for
the KHA1000.
-134 Suffix too long
The suffix contained more than twelve characters.
-138 Suffix not allowed
A suffix was encountered after a numeric element which does not
allow suffixes.
-140 Character data error
Generated when parsing a character data element.
-141 Invalid character data
Either the character data element contains an invalid character, or the
element is not valid.
-144 Character data too Long
The character data element contains more than twelve characters.
-148 Character data not allowed
A legal character data element was encountered where prohibited by
the KHA1000.
-150 String data error
Generated when parsing a string data element.
-160 Block data error
Generated when parsing a block data element.
-170 Expression error
Generated when parsing an expression data element.
-180 Macro error
Generated when defining a macro or executing a macro.
Содержание KHA1000
Страница 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Страница 8: ...vi KHA1000 ...
Страница 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 22: ...xx KHA1000 ...
Страница 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Страница 30: ...1 8 KHA1000 ...
Страница 68: ...3 20 KHA1000 ...
Страница 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Страница 156: ...5 54 KHA1000 ...
Страница 183: ...Other Measurements This chapter explains measurements for other than harmonic current and voltage fluctuation tests 7 ...
Страница 193: ...Remote Control This chapter explains how to connect the remote interface and how to control it using SCPI commands 8 ...
Страница 260: ...9 6 KHA1000 ...
Страница 261: ...Specifications This chapter explains the specifications of this product 10 ...
Страница 272: ...I 4 KHA1000 ...
Страница 273: ......