8-16
KHA1000
8.6
SCPI and IEEE-488.2 Common Commands
*CLS
Clears all event registers including the status bytes, event status, and error queue.
Cancels the completion wait operation by *OPC/*OPC?
Command
*CLS
*ESE
Sets the event status enable register calculated by the event summary bit (ESB) of
the status byte.
Command
*ESE <NR1>
*ESE?
Parameter
Value:
0 to 255.
An SCPI error (-222, “Data out of range”)
occurs if outside the range.
Example:
Sending *ESE16 sets bit 4 of the event status enable register. Every
time the execution error bit (bit 4) of the event status register is set, the
summary bit (ESB) of the status byte is set.
Response
Returns the value of the event status enable register in the <NR1> format in
response to *ESE?.
*ESR
Queries the event status register.
Command
*ESR?
Response
Returns the value of the event status register in the <NR1> format and clears the
register in response to *ESR?.
See
See
See
Содержание KHA1000
Страница 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Страница 8: ...vi KHA1000 ...
Страница 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 22: ...xx KHA1000 ...
Страница 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Страница 30: ...1 8 KHA1000 ...
Страница 68: ...3 20 KHA1000 ...
Страница 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Страница 156: ...5 54 KHA1000 ...
Страница 183: ...Other Measurements This chapter explains measurements for other than harmonic current and voltage fluctuation tests 7 ...
Страница 193: ...Remote Control This chapter explains how to connect the remote interface and how to control it using SCPI commands 8 ...
Страница 260: ...9 6 KHA1000 ...
Страница 261: ...Specifications This chapter explains the specifications of this product 10 ...
Страница 272: ...I 4 KHA1000 ...
Страница 273: ......