5-22
KHA1000
Menu when Class D is selected
F1
Definition of
Power
Window
Average
Specified
Select the actual input power used to
calculate limit values. Normally, use
“Window” or “Average” Specified values
are not defined in the standard.
•
Window
Maximum value of actual power for
each window
•
Average
Average value of actual power in
entire test period
•
Specified
Specified value for manufacturers
F2
Specified Power
(W)
Numeric
value
Selecting Specified of the PF & Fund
Curr lets you enter a value. The entry
range is 0 W to 4 000 W.
F3
Ignore
75 W
Enabled
Disabled