![Keysight Technologies N5171B EXG Скачать руководство пользователя страница 115](http://html1.mh-extra.com/html/keysight-technologies/n5171b-exg/n5171b-exg_user-manual_1962194115.webp)
Keysight EXG and MXG X-Series Signal Generators User’s Guide
99
Optimize Performance
Using External Leveling (N5173B/83B Only)
4.
Determine the output amplitude range, see
“Determining the Signal Generator’s Amplitude
5.
Set the displayed power value, see
“Adjusting the Signal Generator Display’s Amplitude
Equipment Setup
Set up the equipment as shown in
. Place the external detector (detector
and coupler/power splitter) as close as possible to the DUT.
Recommended Equipment
— Keysight 8474E negative detector
— Keysight 87301D directional coupler
— cables and adapters, as required
Figure 4-22
Typical External Leveling Setup using a Directional Coupler
Configuring the Carrier
1.
Press
Preset
.
2.
Set the carrier frequency.
3.
Set the power level to 0 dBm:
—
If the signal generator has Options 1E1 and 520, set the output attenuator to zero dBm:
a.
Press
AMPTD
>
Atten/ALC Control
>
Atten Hold Off On
to On.
b.
Press
Set Atten
>
0
>
dB
.
c.
Press
Set ALC Level
>
0
>
dBm
.
Selecting External Leveling
Press
AMPTD
>
Leveling Control
>
Leveling Mode
>
Pwr Meter Cont
.
Leveled Signal
Signal Generator
Coupler
Amplifier
Negative Detector
RF OUTPUT
ALC INPUT
Содержание N5171B EXG
Страница 1: ...User s Guide Keysight X Series Signal Generators N5171B 72B 73B EXG N5181B 82B 83B MXG...
Страница 4: ...iv...
Страница 356: ...340 Keysight EXG and MXG X Series Signal Generators User s Guide BERT Option UN7 Verifying BERT Operation...
Страница 452: ...436 Keysight EXG and MXG X Series Signal Generators User s Guide Troubleshooting Contacting Keysight Technologies...