
420
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Multitone and Two-Tone Waveforms (Option 430)
Using Multitone Modulation
2.
Set the output amplitude to 0 dBm.
3.
Press
RF On/Off
.
The multitone waveform is now available at the signal generator’s RF OUTPUT connector.
Applying Changes to an Active Multitone Signal
If the multitone generator is currently in use (
Multitone Off On
set to On) while changes are made
in the
Multitone Setup
table editor, you must
apply
the changes before the updated waveform
will be generated.
From the
Multitone Setup
table editor, press the following key to apply the changes and
generate a multitone waveform based on the updated values:
Apply Multitone
Storing a Multitone Waveform
In this example, you learn how to store a multitone waveform. If you have not created a multitone
waveform, complete the steps in the previous section,
Creating a Custom Multitone Waveform
1.
Press
Load/Store
>
Store To File
.
If there is already a file name from the
Catalog of MTONE Files
occupying the active entry
area, press the following keys (see
):
Edit Keys
>
Clear Text
2.
Enter a file name (for example,
5TONE
) using the alpha keys and the numeric keypad with a
maximum length of 23 characters (see
3.
Press
Enter
.
The multitone waveform is now stored in the
Catalog of MTONE Files
.
The RF output amplitude, frequency, and operating state settings are not
stored as part of a multitone waveform file. Similarly, the multitone
settings are not stored as part of the instrument state. Therefore, in most
cases you should save both the instrument states and the multitone
settings to be able to restore all of your settings later.
Содержание N5171B EXG
Страница 1: ...User s Guide Keysight X Series Signal Generators N5171B 72B 73B EXG N5181B 82B 83B MXG...
Страница 4: ...iv...
Страница 356: ...340 Keysight EXG and MXG X Series Signal Generators User s Guide BERT Option UN7 Verifying BERT Operation...
Страница 452: ...436 Keysight EXG and MXG X Series Signal Generators User s Guide Troubleshooting Contacting Keysight Technologies...