76
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Optimize Performance
Using the Dual Power Meter Display
17.
Press
Done
Calibration progress bar is displayed. Refer to
.
Figure 4-12
Running Calibration(s) Bar (Calibrating Sensor)
18.
Press
Return
>
Channel B
to On
19.
The current power meter sensor reading should be displayed on the signal generator in the
ChB portion of the display and in the upper right corner of the display under Power Meter and
to the left of the Power Meter power sensor reading.
Figure 4-13
Channel B Power Sensor Displayed on MXG/EXG
20.
The power sensors are now ready to be connected in a measurement setup.
For details on
each key, use key
help as described
on
.
For details on
each key, use key
help as described
on
.